It is mostly used for WHO or UN aid project, the plunger of syringe was locked to prevent the reuse, help eliminate touch contamination.
In all cases, once the auto-disable feature has been activated, It was not possible to re-use the syringe and the needle under the normal conditions of use.
It was not be possible to override the autodisable feature when tested in accordance with the test method of
ISO7886-3 Annex C.
If an integrated needle is used, it shall become an integral part of the syringe and can not be detached.
The auto disable feature is automatically and remains effective from the point when 50% of the intended fixed dose has been delivered Invention patented.
Code No | Size | Needle Size |
ADS1-0.05 | 0.05ml | 23G-27G |
ADS1-0.1 | 0.1ml | 23G-27G |
ADS1-0.2 | 0.2ML | 23G-27G |
ADS1-0.3 | 0.3ML | 23G-27G |
ADS1-0.4 | 0.4ML | 23G-27G |
ADS1-0.5 | 0.5ML | 23G-27G |
ADS1-1 | 1ML | 23G-27G |
Syringe with non-integrated needle shall have a minimum needle unicon force applied as push or pull in the direction of the needle axis in accordance with ISO 7864:1993.
The autodisable feature is automatically activated on completion of the delivery of the intended fixed dose.
Central luer slip or eccentric luer slip.
Code No | Size |
ADS2-2 | 2ml |
ADS2-3 | 3ml |
ADS2-5 | 5ml |
ADS2-10 | 10ml |